A spectrometer turns light into information: from the wavelength you can recognize a substance, assess the condition of a plant, or detect a contaminant. The smaller, more portable, and cheaper the instrument is meant to be, the harder it is to maintain spectral quality, and the less predictable the operating conditions, the more important a well-thought-out optical design becomes.
In this article we show how to approach the design of a spectrometer or a multispectral camera for environmental and agricultural applications, and how TracePro and OSLO from Lambda Research support such a project. A note on sources: the manufacturer describes spectrometer applications in environmental monitoring and remote sensing, whereas we treat agriculture here as an application area that follows from the physics of the measurement, not as an example named by Lambda Research.
Why spectroscopy goes into the field
Devices that measure in the field, on a farm, in water, or from the deck of a vehicle or aircraft require a different approach than a laboratory instrument. Lambda Research mentions airborne and portable systems as a typical area of application for hyperspectral and multispectral imaging.
Such a device has several common features:
- a compact housing in which the elements lie close to each other,
- operation in variable lighting and temperature,
- a need for high sensitivity with small aperture sizes,
- a requirement for robustness and repeatability of measurements.
Each of these has optical consequences that are easier to study on a model than on successive prototypes.
Which spectrometers can be modeled
In its description of TracePro applications in spectroscopy, Lambda Research lists, among others:
- Raman spectroscopy,
- absorption and emission spectroscopy,
- Fourier transform spectroscopy (FTS),
- fluorescence spectroscopy,
- hyperspectral and multispectral imaging,
- grating and interferometric spectrometers,
- monochromators.
This is a wide range, but the common denominator is the same: light of many wavelengths passes through a system of elements that are meant to separate or filter it, and the detector is meant to measure the result without disturbance.
TracePro in a spectrometer project
TracePro uses non-sequential Monte Carlo ray tracing for many wavelengths and lets you build a 3D model of the whole instrument with its geometry, sources, materials, surfaces, and detectors. In the context of spectrometers, Lambda Research points to, among others:
- modeling of diffraction gratings, lenses, mirrors, and filters,
- analysis of beam splitters and prisms,
- modeling of diffractive elements, including holographic elements and computer-generated holograms, with the phase described by Zernike polynomials,
- libraries of wavelength-dependent material and surface properties,
- import of geometry from CAD systems,
- photorealistic visualization of the model.
For diffraction gratings, the ability to determine the energy efficiency for different angles of incidence and to calculate the flux in individual orders is important, because it translates into how much light actually goes into the useful diffraction order and how much into the unwanted ones.
Stray light and spectral crosstalk
In a spectrometer, the error does not always result from an inaccurate optical system. It is often caused by reflections and scattering that direct light of the wrong wavelength onto the detector. Lambda Research lists stray light, ghost reflections, and spectral crosstalk as the main sources of measurement error in hyperspectral systems, as well as multiple reflections and scattering that degrade spectral purity.
TracePro offers a Stray Light Analyzer for identifying reflections and unwanted light paths. In a compact instrument this lets you assess which housing or mount surfaces contribute the most, and check a change of material, geometry, or finish before a prototype is made. The basics are covered in the article Stray light analysis: what it is.
OSLO: the imaging and collimating system
Before analyzing the whole instrument you have to design its optical system, for example the system that collimates light before the grating and the system that focuses it onto the detector. OSLO is used to design and optimize such systems and to analyze aberrations and tolerances. The OSLO model can then be transferred to TracePro to check it together with the grating, housing, and detector. We write about collimators in the article Collimator design.
Example workflow for designing a spectrometer
1. Define the spectral range and resolution
Determine the wavelength range, the required spectral resolution and sensitivity, and the operating conditions of the instrument.
2. Design the optical system in OSLO
Optimize the collimating and focusing system and check the aberrations across the whole wavelength range.
3. Choose the dispersive element
Select a grating or another element that separates light and define its parameters.
4. Build a 3D model in TracePro
Add the grating, mounts, housing, slit, and detector, and define wavelength-dependent surface properties.
5. Assess the energy distribution among diffraction orders
Check how much light goes into the useful order and how much into the unwanted ones.
6. Analyze stray light
Identify the unwanted light paths and introduce corrections such as baffles, a change of finish, or a change of geometry.
7. Check the tolerances
Assess the effect of manufacturing and assembly deviations on spectral quality.
8. Correlate the model with measurements
Compare the spectra of reference standards with the model and adjust the simulation parameters.
Which tool for which task
| Task | Tool |
|---|---|
| Design of the collimating and focusing system | OSLO |
| Aberration and tolerance analysis | OSLO |
| 3D model of the spectrometer with housing | TracePro |
| Diffraction gratings and diffraction orders | TracePro |
| Stray light and spectral crosstalk | TracePro |
| Coatings, filters, and λ-dependent materials | TracePro |
The most common mistakes in spectrometer design
Analyzing only one wavelength
A spectrometer works across a range of the spectrum, so the simulation should cover many wavelengths, not just one.
Ignoring stray light
In a compact housing, reflections and scattering can dominate the signal.
Ideal surfaces
Real materials and finishes scatter and reflect light, and the model should reflect this.
No analysis of diffraction orders
Energy in unwanted orders can reach the detector and distort the spectrum.
Ignoring field conditions
A field device works in variable temperature and lighting, which has to be taken into account in the design and the tolerances.
Summary
Spectrometers and multispectral systems for environmental monitoring and agriculture have to combine small size with a clean spectral signal. Achieving this requires a model that covers not only the optical system but also the grating, housing, coatings, and stray light.
OSLO lets you design the optical system and check its tolerances, and TracePro carries it over to a 3D model of the instrument in which you can analyze gratings, diffraction orders, reflections, and spectral crosstalk. This way, design decisions can be made on the basis of the model, before an expensive prototype is built.
Manufacturer resources
Applications of TracePro in spectroscopy: Lambda Research: Spectroscopy Simulation Using TracePro
Hyperspectral and multispectral imaging and stray light control: Lambda Research: Hyperspectral and Multispectral Imaging Systems
Stray light analysis in TracePro: Lambda Research: Stray Light Analysis with TracePro