An endoscope, an optical diagnostic system, or a surgical light are devices in which the optics must work in a very limited space, with strictly defined illumination, and with no room for compromise in image quality. That is why medical optics design relies on two levels of simulation: the design of the imaging system, and the analysis of the whole device together with its illumination and mechanics.

Optics in medicine: simulating endoscopes and optical diagnostic systems in OSLO and TracePro
Request a quote Start a trial

Medical optics combines two kinds of requirements that occur together less often in typical industrial devices. First, the imaging system must deliver high image quality despite its small size. Second, light must be delivered to the observed site in a controlled way, without reflections and scattering that would degrade the image.

That is why the design of such a device is best carried out on two levels: from the imaging optics in OSLO to the model of the whole device in TracePro. Both programs come from Lambda Research and are designed to work together.

Two levels of the problem: image and light

In an endoscope or a diagnostic device you can distinguish two light paths. The first is the imaging path, that is, the path of light from the observed object through the objective to the detector. The second is the illumination path, that is, the path of light from the source to the object.

The two paths are linked. A good objective will not help if the illumination is uneven or if reflections from the mount elements reach the detector as stray light. Conversely, well-designed illumination will not rescue the image if the imaging system has excessive aberrations across the field of view.

OSLO: designing the imaging system

OSLO (Optics Software for Layout and Optimization) is a program specialized in the design and optimization of lens systems. It supports geometric and wave analysis and aberration correction. In a medical optics project this means, among other things:

  • selecting the lens system and optimizing it against the adopted image quality criteria,
  • analyzing aberrations across the whole field of view,
  • assessing the behavior of the system at different wavelengths,
  • analyzing manufacturing and assembly tolerances.

Tolerance analysis is particularly important. Miniature optics are sensitive to manufacturing deviations, and medical devices are produced in series. We describe this in more detail in the article Tolerance analysis of optical systems in OSLO.

TracePro: illumination, mechanics, and stray light

TracePro uses non-sequential Monte Carlo ray tracing and lets you build a 3D model of the whole device: light sources, the optical system, mounts, the housing, and the detector. Lambda Research notes in its materials that in endoscopes and surgical lighting systems TracePro helps simulate light paths and energy delivery.

In practice this makes it possible to answer questions that the design of the objective alone cannot resolve:

  • whether the illumination is distributed as intended,
  • what fraction of the light actually reaches the observed site,
  • whether reflections from mounts or the housing reach the detector,
  • how materials and surfaces affect the light distribution,
  • which unwanted light paths need to be limited.

Among the applications of TracePro in the biomedical area, Lambda Research also lists biosensors, molecular spectroscopy, laser scanning microscopy, and medical imaging.

From the OSLO model to the device model

The two programs form a consistent workflow. A lens system designed and optimized in OSLO can be exported, and TracePro accepts the LEN file format of OSLO. In TracePro, the optical model is reconstructed in a 3D environment in which you can add real light sources, mechanical elements, baffles, and the housing.

This turns the question "does the objective meet the specification?" into "does the complete device meet the specification?". Lambda Research describes such a workflow using the example of a lidar receiver, in which, after the lens system was optimized in OSLO, TracePro was used to model the complete assembly with its housing and internal coatings and to detect stray light problems. The same logic applies to medical optics.

Stray light in small devices

In miniature devices the distances between elements are small and mechanical surfaces lie close to the optical path. This favors reflections and scattering that degrade image contrast. We describe this phenomenon in the article Stray light analysis: what it is.

TracePro lets you identify the unwanted light paths, assess the contribution of individual surfaces, and check the effect of design changes, for example a change of material or surface finish, before a prototype is made.

Example workflow for designing a device with medical optics

1. Define the requirements

Determine the field of view, the resolution, the wavelength range, the device dimensions, and the required illumination level of the observed area.

2. Design the imaging system in OSLO

Optimize the lens system and check the aberrations across the whole field of view.

3. Perform a tolerance analysis

Check how manufacturing and assembly deviations affect image quality, and identify the critical parameters.

4. Build the device model in TracePro

Import the system from OSLO, add the light source, mounts, housing, and detector, and assign surface and material properties.

5. Analyze the illumination

Check the light distribution in the observed area and how much energy reaches the target.

6. Analyze stray light

Identify the unwanted light paths and introduce design corrections.

7. Correlate the model with prototype measurements

Compare the simulation results with measurements and, if needed, update the model parameters.

Which tool for which task

Task Tool
Design and optimization of the objective OSLO
Aberration analysis across the field of view OSLO
Manufacturing and assembly tolerance analysis OSLO
3D model of the device with mounts and housing TracePro
Illumination and energy delivery analysis TracePro
Stray light and unwanted reflections TracePro
Biosensors, spectroscopy, microscopy TracePro

The most common mistakes in medical optics simulation

Designing the objective alone, without the illumination

Image quality depends on the imaging system and on the illumination. Both should be analyzed together.

Neglecting the mechanics

Mounts, the housing, and positioning elements can reflect and scatter light, so they should be part of the model.

Analyzing only the nominal system

In miniature optics, manufacturing deviations have a large effect on the result. Tolerances should be checked before the prototype.

Ideal surfaces

Real materials and surface finishes scatter and reflect light. The model should reflect their properties.

Treating simulation as validation

Simulation supports the design but does not replace measurements and the procedures required for medical devices.

Summary

Medical optics design requires combining two perspectives: image quality and control of light in the whole device. OSLO lets you design and optimize the imaging system and check its tolerances. TracePro carries that system over to a 3D model of the device, in which you can analyze illumination, mechanics, and stray light.

This way the engineer can detect early the problems that in a typical process would only appear on the prototype, and make design decisions based on the model rather than on measurements alone.

Manufacturer resources

Description of the cooperation of OSLO and TracePro in an optical system project: Lambda Research: Optical System Design with TracePro and OSLO

Workflow from OSLO to TracePro: Lambda Research: Integrating TracePro and OSLO for Comprehensive Optical System Design

Frequently asked questions
Optics in medicine: simulating endoscopes and optical diagnostic systems in OSLO and TracePro

What software is suitable for designing endoscopes?

For the design of the imaging system itself, that is, the objective, OSLO is suitable, a program for sequential design and optimization of optical systems. For analyzing the whole device, including illumination, mounts, and stray light, use TracePro, a program based on non-sequential Monte Carlo ray tracing.

Is TracePro used in medical optics?

Yes. Lambda Research lists medical imaging and endoscopy, biosensors, molecular spectroscopy, and laser scanning microscopy among the applications of TracePro. In the manufacturer's materials, endoscopes and surgical lighting systems appear as examples of devices in which TracePro helps simulate light paths and energy delivery.

What is the difference between the roles of OSLO and TracePro in an endoscope project?

OSLO is used to design and optimize the lens system and to analyze aberrations and tolerances. TracePro lets you place that system in a 3D model of the device together with the light source, the mounts, and the detector, and check how light behaves in the real construction.

Can an OSLO model be transferred to TracePro?

Yes. A lens system model made in OSLO can be exported and opened in TracePro, which accepts the LEN file format of OSLO. In TracePro you can then add light sources, mechanical elements, baffles, and the housing.

Why is tolerance analysis important in medical devices?

Miniature optics are especially sensitive to manufacturing and assembly deviations, and the devices are produced in series. Tolerance analysis in OSLO lets you check how parameter deviations will affect image quality and determine which of them are critical, before a prototype is built.

Does simulation replace the validation of a medical device?

No. Simulation supports design and lets you detect optical problems early, but it does not replace prototype measurements, studies, or the procedures required for medical devices. It should be treated as a tool for design decisions, and its results are worth correlating with measurements.